Maintenance
D A N G E R
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
D A N G E R
Electrocution hazard. Remove power from the instrument before doing maintenance or service
activities.
Maintenance schedule
Table 15
shows the recommended schedule of maintenance tasks. Facility requirements and
operating conditions may increase the frequency of some tasks.
Table 15 Maintenance schedule
Task 6 months As necessary
Clean the instrument on page 40 X
Clean leaks and spills on page 40 X
Replace the water filter
1
X
Replace the UV lamps
1
X
Replace a fuse on page 41 X
1
Refer to the instructions supplied with the replacement part.
Clean the instrument
Clean the exterior of the instrument with a moist cloth and a mild soap solution.
Clean leaks and spills
C A U T I O N
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local, regional and
national regulations.
W A R N I N G
Personal injury hazard. The sample line contains water under high water pressure that can burn skin if
hot. Qualified personnel must remove the water pressure and wear personal protective equipment
during this procedure.
N O T I C E
Tighten a new ferrule 1¼ turns the first time it is tightened. Then, only tighten the ferrule 1/8 turn to make a
connection. Do not tighten the fittings more as this may cause damage to the fittings or ferrules and cause leaks.
If a leak occurs in the tubing, the sample water will drain from small holes in the bottom of the
enclosure.
For leaks caused by a valve, worn tubes or other parts, send the instrument to the manufacturer for
service. Refer to Send the instrument for service on page 42
.
40 English