Table 18 Warning codes (continued)
Code Message Description
SC153–SC156 BottleXMarkedAsUsed The bottle installed in location is marked as used.
Replace the bottle with the correct standard.
SC160 ExcursionBottle4NotEmpty The excursion bottle is not empty. Replace the
excursion bottle with an empty bottle.
SC165 LampSwitch One lamp is weak or has failed. Replace the lamp.
SC172 RougeDetected One or more of the 500 ppb standards took longer than
expected to oxidize. Clean the cell.
SC178 ExcursionCaptured An excursion sample was captured in an excursion
bottle. Remove the sample from the analyzer.
SC180 ConductivityMeterAlarm The analyzer has not passed the conductivity meter
test.
SC181 BottleTestCancelled The bottle test was cancelled by the user. Start the test
again.
SC182–SC185 BottleXExpired The bottle installed in the location has expired. Replace
the bottle with the correct standard.
SC186 BottleTestCanceled-ExpiredStandard The bottle test was canceled because there is an
expired standard bottle in the bottle bay.
SC189 IncorrectRFID The standards bottle has a incompatible RFID tag.
SC192 ExcursionNotConfirmed The excursion is not confirmed.
Measurement alarm codes
Measurement alarms occur when TOC or conductivity levels in the process water are more than the
set limits. Refer to Table 19.
The digital outputs that identify a TOC alarm or conductivity alarm change their state when a
measurement alarm is present.
Refer to Configure the digital output alarms on page 31. The three
analog outputs continue to represent the last measured value when a measurement alarm is present.
Table 19 Measurement alarm conditions
Code Message Description
SC161 TOCAlarm The measured TOC is more than the user-defined TOC level.
SC162 ConductivityAlarm The measured conductivity is more than the USP645 conductivity level.
Diagnostics
Review the data log usage and firmware version
1. Push << to show the sliding menu.
2. Push DIAGNOSTICS, then select the General Tab.
The data log percent usage and the firmware version show.
Review the UV lamp information
See the lamp status and the number of operational hours until the next lamp replacement.
46 English