Specifications
Specifications are subject to change without notice.
Specification Details
Dimensions (W x D x H) 59.7 x 22.9 x 25.4 cm (23.5 x 9 x 10 in.)
Enclosure Quick-connect wiring model: IP46
Conduit wiring model: IP56
Weight 13.6 kg (30 lb)
Pollution degree 2
Installation category II
Protection class I
Power requirements 100/230 VAC, ±10%, 65 W, 50/60 Hz
Altitude 4000 m (13,125 ft) maximum
Operating temperature 10 to 40 °C (50 to 104 °F)
Humidity 5–95%, non-condensing
Sample temperature 1 to 95 °C (34 to 203 °F)
Sample flow rate 60 to 300 mL/min
Sample pressure 0.70 to 7.0 kg/cm² (10 to 100 psi)
Temperature accuracy ±0.5 °C
UV lamps 2x, with UV Detect technology
User interface Color touch screen
TOC Range: 0.5 to 2000 ppb as carbon
Accuracy: ±1 ppb or ±5%, whichever is greater
Repeatability: ±0.3 ppb or ±1%, whichever is greater
Detection limit: 0.5 ppb
Conductivity Range for conductivity mode: 0.05 to 150 μS/cm at 25 °C
Accuracy: ±2% over the full range
Range for TOC mode: 0.05–5.0 µS/cm
Outputs Analog: Three isolated
Discrete: Four isolated, configurable
Inputs Discrete: Two isolated
Certifications CE Certified. Listed to UL and CSA safety standards by ETL.
Warranty 1 year
General information
In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential
damages resulting from any defect or omission in this manual. The manufacturer reserves the right to
make changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer’s website.
English 5