Validate the pump operation
Use the pump test icon to test and/or calibrate the pumps.
1. Push << to show the sliding menu.
2. Push DIAGNOSTICS, then select the Tests tab.
3. Push PUMP TEST.
4. Put a 50-mL graduated cylinder at the water outlet of the analyzer. Do not use tubing.
5. Make sure that a bottle is installed in position 2 to enable the pump to be primed.
6. Select an option.
Option Description
Calibrate Pump Calibrate the pump.
Start/Stop Pump Start or stop the pump. Start Pump—Open the pump valve and the grab sample bottle
valve, then set the pump to on. The pump stays on until Stop Pump is pushed. Stop
Pump—Set the pump to off and close both valves.
Prime Pump Open the pump valve and the grab sample valve, then set the pump to on. The pump
operates long enough to prime. Then, the pump switches off and both valves close.
Load Bottle Install bottles in the bottle bay.
Unload Bottle Remove bottles from the bottle bay.
7. Push START.
If Calibrate Pump was selected, the analyzer pumps water for 30 seconds.
8. For pump calibration, enter the volume pumped as measured in the graduated cylinder.
The test results give the:
• Test Time— Always 30 seconds
• Expected Results—The calculated amount of water that should be measured during the test.
The formula is: V = Flow rate ÷ Test time, where Flow Rate = The target flow rate of the pump
as entered in the Actual Results text box.
• Actual Results—The actual amount of water measured during the test.
9. Push the previous icon to save the calibration results.
Do an RFID test
Use the RFID test to test the RFID reader system.
1. Push << to show the sliding menu.
2. Push DIAGNOSTICS, then select the Tests tab.
3. Push RFID TEST.
4. Install four standards bottles equipped with RFID tags in the bottle bay. The bottle can be new or
used standards bottles.
5. Push START.
The system automatically senses those locations where RFID tags are present.
6. To cause a test failure, push the unlock icon during the test to open the bottle bay door. The
system will return a failed test because the bottles are no longer positioned properly with the door
open.
Do a fan test
Use the fan test to test the analyzer fans. The analyzer contains an internal online heat exchanger
that operates to maintain proper water temperature to the analysis cell. Two fans cool the heat
exchanger.
1. Push << to show the sliding menu.
2. Push DIAGNOSTICS, then select the Tests tab.
English 49