Option Description
Standby Put the instrument in standby mode (valve open)
Offline Put the instrument in offline mode (valve closed)
TOC manual sample Manual TOC measurement for online sample
Self clean Configure and start a clean cycle
4. Push ENTER.
Calibration
C A U T I O N
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets
(MSDS/SDS) for safety protocols.
TOC calibration
Calibrate the instrument for TOC measurements with default settings or with custom settings. Use
default settings with packaged standards kits that use RFID tags on the bottles. Use custom settings
for less replicates or less bottles.
1. Make sure that the TOC standards are at room temperature.
2. Push << to show the sliding menu.
3. Push BOTTLE MODE, then RUN STANDARDS.
4. Push TOC CALIBRATION.
5. Select default or custom settings.
6. Obey the screen prompts to set the number of repetitions and confirm the sucrose standard
values.
Note: To save the new settings as the default, push SAVE AS DEFAULT.
7. Install the standards in the order shown on the screen. If the bottle bay door closes too soon,
push UNLOCK to open the door.
8. Push START.
9. When the calibration is complete, the results are shown. If a printer is attached, the report can be
sent to a printer. Select the necessary action to complete the calibration.
Push CHECK to accept the calibration and to remove the bottles. This option will not show if
the Autoaccept setting is enabled.
Push CANCEL to repeat the calibration and discard the results.
Push EXPORT to send the data to the supplied USB flash drive.
Conductivity calibration
Use the optional calibration resistor for this procedure. Figure 11 shows the location of the calibration
resistor.
1. Make sure that the conductivity standards are at room temperature.
2. Push << to show the sliding menu.
3. Push BOTTLE MODE, then RUN STANDARDS.
4. Push CONDUCTIVITY CALIBRATION.
5. Obey the screen prompts.