Table 16 Error codes (continued)
Code Message Description
SC151 LampExtinct The current lamp has failed. Replace the lamp.
SC163 UnableTo-StopCurrentOperation The analyzer can not stop its current operation.
SC164 UnknownError The analyzer has experienced an unknown error.
SC167 WatchDogIoCard The I/O card has stopped responding.
SC168 WatchDogTocAlgoithm The TOC algorithm has stopped responding.
SC169 WatchDogPrinter The printer has stopped responding.
SC170 WatchDogRs232 The RS-232 module has stopped responding.
SC177 Exception An exception was thrown.
SC188 TOCOverLimit The TOC measurement is more than the instrument range.
Critical error codes
In a critical error condition, the instrument cannot function. The error shows as a full-screen flashing
message. The digital output that represents the error condition and the three analog outputs (2 or
22 mA) change their state to identify when an error condition is present. Refer to Alarm indication
on page
42.
Operation stops until the error condition is corrected. Touch the screen interface to accept the alarm
and see more information. Refer to Table 17.
Table 17 Critical error codes
Code Message Description
SC110 UnableTo-StartIoModule The analyzer can not start the IO module.
SC111 UnableTo-StartTocModule The analyzer can not start the IO module.
SC114 UnableTo-StartEthernetModule The analyzer can not start the Ethernet module.
SC152 LampUnableToSwitch The analyzer can not switch from one lamp to the other.
SC173 RangeTable-FailedToLoad The RangeTable.txt file could not be loaded.
SC174 InstrumentTemperature The ambient temperature is more than the instrument specifications.
Decrease the instrument and environmental temperature.
SC175 BothLampsBad Both UV lamps have failed. Replace both lamps.
SC176 NoCond-TempReadings The analyzer can not measure conductivity and temperature. Restart
the analyzer.
SC179 HardwareInitializationFailure An error occurred during the hardware initialization.
Warning codes
When a warning condition is present, a warning indicator (flashing red rectangle on the display)
shows on the display to identify that a warning message is pending in the alarm log. The digital
output that represents an error condition changes its state to identify when a warning condition is
present. The three analog outputs do not identify when a warning condition is present. Refer to Alarm
indication on page
42.