Operation may continue when a warning condition is present. Refer to Table 18.
Table 18 Warning codes
Code Message Description
SC100–SC103 IncorrectBottle-LoadedinPositionX Incorrect bottle detected in position X. The analyzer can
not do the bottle test operation. Examine and change
the bottle as necessary.
SC104–SC107 NotEnough-FluidinBottleX There is not enough fluid in bottle to complete the
operation. The analyzer aborts operation and returns to
its normal operating mode.
SC108 UnableTo-ReadSettings Some or all of the settings could not be read from the
settings file.
SC109 UnableTo-WritetoSettings The analyzer can not write to the settings file.
SC118 UnableTo-ChangeLanguageSettings The analyzer can not change the language settings.
SC119 UnableTo-ReadLanguageSettings The analyzer can not read the language settings
SC120 UnableTo-ChangeEthernetSettings The analyzer can not change the Ethernet settings.
SC121 UnableTo-ReadEthernetSettings The analyzer can not read the Ethernet settings.
SC122 UnableTo-PrintSystemSettings The analyzer can not print the system settings.
SC125 UnableTo-StartLampTest The analyzer can not start the lamp test.
SC126 UnableTo-StopLampTest The analyzer can not stop the lamp test.
SC127 UvLampLifeExceeded The maximum lamp hours of usage have been
exceeded or the UV Detect
™
output is below the
acceptable limit. The lamp has been set to a status of
marginal. Replace the lamp.
SC129 UnableTo-
ReadInformationFromBottle
The analyzer can not read the information from the
bottle.
SC130 UnableTo-WriteInformationToBottle The analyzer can not write the information to the bottle.
SC132 UnableTo-StartPumpTest The analyzer can not start the pump test.
SC133 UnableTo-StopPumpTest The analyzer can not stop the pump test.
SC134 UnableTo-CalibratePump The analyzer can not calibrate the pump.
SC135 UnableTo-ExportDataLog The analyzer can not export the data log.
SC136 UnableTo-ExportAuditTrail The analyzer can not export the audit trail.
SC137 UnableTo-AddUser The analyzer can not add a user.
SC138 UnableTo-EditUser The analyzer can not edit a user.
SC139 UnableTo-DeleteUser The analyzer can not delete a user.
SC141 UnableTo-AcknowledgeAlarms The analyzer can not acknowledge alarm.
SC142 UnableTo-SilenceAlarms The analyzer can not delete alarms.
SC143 UnableTo-StartSelfClean The analyzer can not start the self clean.
SC144 UnableTo-StopSelfClean The analyzer can not stop the self clean.
SC145 UnableTo-StartOnlineManualSample The analyzer can not start the online manual sample.
SC147 UnableTo-RestoreFactoryDefaults The analyzer can not restore the factory defaults.
SC150 UnableTo-PrintDataLog The analyzer can not print the data log.