Configure the analog outputs
Set the parameter (TOC, conductivity or temperature), upper limit and lower limit for each analog
output. Set the value at the analog outputs when an error occurs.
1. Push << to show the sliding menu.
2. Push the SETUP icon.
3. Push ANALOGS.
4. Select the General tab.
5. Set the zero scale and full scale values for each parameter (TOC, conductivity and temperature).
Zero scale—The lower limit parameter value that will be represented as 4 mA at the analog
output. Options: TOC = 0 to 1999 ppb, Conductivity = 0 to 19 µS/cm, Temperature = 0 to 99 °C
Full scale—The upper limit parameter value that will be represented as 20 mA at the analog
output. Options: TOC = 1 to 2000 ppb, Conductivity = 1 to 20 µS/cm, Temperature = 1 to
100 °C
6. Set the value at the analog outputs that will be present when an error occurs. Options: 2 mA
(default), last output or 22 mA
7. To enable analog output during bottle mode, select "Analog Outputs during Bottle Mode".
Configure the printer settings
Configure the analysis result printing and the printing frequency for the online TOC mode and
conductivity mode.
1. Push << to show the sliding menu.
2. Push the SETUP icon.
3. Push PRINTER.
4. Select the General tab.
5. Select the type of printer.
6. Select the TOC tab.
7. Select an option.
Option Description
Disabled Disable the printing of online TOC readings.
Continuous Enable continuous printing. A TOC printout occurs at the end of each analysis cycle.
Percent Change Enable a printout to occur only when the change in TOC between successive analyses is
more than the value in the Limit field. Enter a value in the Limit field. Options: 1 (default) to
100%
8. Select the Conductivity tab.
9. Select an option.
Option Description
Disabled Disable the printing of online conductivity readings.
Timed Enable continuous printing. A conductivity printout occurs at the end of each interval.
Enter an interval in the Interval field. Options: 1 minute (default) to 24 hours
Percent Change Enable a printout to occur only when the change in conductivity between successive
analyses is more than the value in the Limit field. Enter a value in the Limit field. Options:
1 (default) to 100%
English 33