EasyManua.ls Logo

Hanna Instruments PCA310 - Calibrate the Measuring Cell; Calibration Date and Factor; Calibration Procedure

Hanna Instruments PCA310
80 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Cal.Value:0.14 mg/L
Factor :0.954
Cal. Date:04/01/20
Reset Cal. Factor
Calibration Blank
The PCA310 - PCA340 analyzers have the possibility to calibrate the measuring cell.
When a new calibration is performed, calibration factor is recalculated and all measurements are
multiplied with it.
CALIBRATION DATE AND FACTOR
The last calibration date can be found in the “Chlorine menu” - “Cal. Measuring Cell” - “Cal. Date”.
Calibration date is in the YY / MM / DD format.
A warning “Cl Calibration Old” is displayed if one month elapsed from the last calibration.
The calibrated date is updated after a new calibration is done.
The calibration factor is displayed in “Chlorine menu” - “Cal. Measuring Cell” - “Factor”.
The default calibration factor is 1.000. Each measurement result is multiplied with calibration factor.
The calibration factor could be reset to 1.000 by activating the “Chlorine menu” - “Cal. Measuring
Cell” - “Reset Cal. Factor” function.
CALIBRATION PROCEDURE
To calibrate the measuring cell, follow the steps:
• Withdraw a sample of the measured liquid directly from the drain port of the measuring cell (#12)
by opening its valve (#11) - see figure on page 11.
Note: Withdraw the sample just before the electrovalve stops the liquid flow to the measuring cell.
CALIBRATE THE MEASURING CELL
• With a calibrated meter take a measure of the sample. This is the calibration value.
• Wait for the PCA to display the new reading.
• Go in “Chlorine menu” - “Cal. Measuring Cell” and edit “Cal. Value” field.
• Enter the calibration value and save with “CFM”.
• The calibration coefficient and the calibration date will be updated.
44

Table of Contents

Related product manuals