--Alarms SMS--
Cl Events :Inactive
pH Events :Inactive
ORP Events:Inactive
Temp.Event:Inactive
--Info SMS --
Set :Inactive
Errors :Inactive
SETTING SMS FEATURE
In order to avoid the frequent sending of the SMSs or rapid consuming of the GSM credit when one
measuring channel not work correctly, the SMS sending could be customized.
The options are available in the “General Menu” - “SMS Settings”.
To select the events that generate alarm SMS, set the “Cl Events”, “pH Events”, “ORP Events” or
“Temp. Event” as Active or Inactive. When one of them is active the corresponding alarms and errors
will generate an alarm SMS. When alarm SMS is sent, all current alarms and errors are sent, even if
are not the ones that initiate the SMS.
To customize the info SMS, two options are available:
“Errors” - when active will include in the info SMS all alarms and errors currently active on the
analyzer.
“Set” - when active will include in the info SMS alarm and dosing setpoints.
Alarm SMS
When errors or alarms appears, alarm SMS is sent to the active phone numbers. Alarm SMS is sent
without special request. The events that generate alarm SMS could be selected as described in Setting
SMS Feature paragraph. Alarm SMS is sent once per chlorine cycle.
At each sending of alarm SMS, after 30 seconds, a phone call is made by the analyzer to the pro-
grammed number(s). This is done because the SMS can be received with a considerable delay due to
network overload, while the phone call takes place immediately and has a long ring which is more
likely to be heard.
GSM
The phone call advises the user that something happened on the PCA analyzer and SMS is going to be
received. Is not necessary to answer the phone call and it is suggested to close it without any answer.
A confirmation of the alarm message reception is waited by the analyzer. The confirmation can be done
by simply calling the PCA analyzer phone number (number of the GSM module). The analyzer will
hang up without answering and will send Info SMS (this confirmation is managed as an information
request; see further on for details). Info SMS don’t need confirmation.
64

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments PCA310 and is the answer not in the manual?

Hanna Instruments PCA310 Specifications

General IconGeneral
BrandHanna Instruments
ModelPCA310
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals