67
Note: This particular warning message does not need confirmation.
In this case the user has to recharge or substitute the SIM card. The sending of the repeated
warning messages will be reset when the expiration date is changed. If the expiration date is
reached without any updating of the expired date, then the message “GSM Card expired” ap-
pears on the display and no more SMSs will be sent by the analyzer until the error is deactivated.
To deactivate this error it is necessary to update the SIM expiration date.
If the user has unlimited credit on the SIM card, the “Chk. Charge” must be set “Inactive”.
In this case the value of remaining messages will not be decremented and no check will be
performed on the SIM card expiration date. Moreover, in the info SMS will not be present the
remaining-messages information.
If a problem related to GSM feature occurs during the normal functioning of the analyzer, the
“GSM Not Respond”, “GSM Init failed” or “GSM Network Error” will be displayed and the
analyzer will try repeatedly to initialize the cellular engine. The error will be deactivated only
after a successful initialization.
GSM
MODEM CONNECTION
Modem connection can be established between PCA310 - PCA340 and a remote computer. The
connection allows the user to interrogate the analyzer, from remote position, about its status and
measurements and to change analyzer parameters. Also the log could be downloaded through remote
connection.
A SIM card able to receive data calls must be used in HI504900 GSM module.
To enable the answer to data request, the item “Remote” from “General Menu” - “Serial & GSM
com.” must be set “Active”.
To make the data transfer between PCA310 - PCA340 and remote PC, HI92500 software must be
installed on the PC, and modem must be connected between PC and phone line.
The baud rate, the password and the RS485 address must the same on PCA and on PC application.
Note: If the PC stops the communication for 4 minutes, the GSM call is interrupted by the
analyzer to free the line.
Remaining:0000
RepeatNo:2
Delay :05 min
Remote :Active

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments PCA310 and is the answer not in the manual?

Hanna Instruments PCA310 Specifications

General IconGeneral
BrandHanna Instruments
ModelPCA310
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals