EasyManuals Logo
Home>Hanna Instruments>Measuring Instruments>PCA310

Hanna Instruments PCA310 User Manual

Hanna Instruments PCA310
80 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #39 background imageLoading...
Page #39 background image
39
Priming in progress
...
168s
The settings related to chlorine measurement are grouped in “Chlorine Menu”. The following options
are available:
REAGENT CHANGING
One set of reagents is enough for at least 16000 samples.
The remaining doses of reagent are displayed on one chlorine measuring panel.
When the reagent is changed, several actions must be performed:
Prepare the reagent and install the new bottles as described in initial preparation and installation
chapter.
Prime the reagent pump if needed or simply reset the reagent counter.
The used reagent doses and the remaining reagent doses could be viewed on the first two lines when
enter in “Chlorine Menu” - “Reagent change”.
If “Chlorine Menu” - “Reagent change” - “Reset Reag. Counter” function is selected and confirmed,
the used doses become 0 and the remaining doses become 16000. This command does not perform
a priming of the reagent pump.
If “Chlorine Menu” - “Reagent change” - “Prime Reag. Circuit” function is selected and confirmed,
the dosing pump is turned on for 180 seconds. The remaining time is displayed on the right-down
corner of the display.
The process could be terminated by pressing ESC at any moment.
CHLORINE SETTINGS

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments PCA310 and is the answer not in the manual?

Hanna Instruments PCA310 Specifications

General IconGeneral
BrandHanna Instruments
ModelPCA310
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals