One or two phone numbers associated with the service - to which the messages will be sent – must
be set. The number has to be entered in the fields named “No1” and “No2”. The space reserved for
a phone number is 15 digits.
The phone numbers have to be inserted using the international format excluding the starting +
character and without any space inside the number.
Example: if the cellular number is +39123456789 (+39 is the country code), the number to be
stored in the PCA310 - PCA340 is 39123456789.
The numbers could be activated or inactivated. The SMS is sent only to active numbers. If both numbers
are inactive, no SMS is sent even if both numbers are correctly entered. To activate the phone numbers,
set “Active” in the “Phone1” and “Phone2” fields.
The SMS sending has to be activated by selecting “Active” in the “Send SMS” field. If “Send SMS”
is set to “Inactive” then no SMS will be sent even if the phone numbers are active.
The number of repeated alarm SMSs could be set in the field “RepeatNo”. The range is between 0
and 5. If RepeatNo is set to 0 than only one SMS is sent.
The delay between two repeated SMSs is set in the “Delay” item. The range is between
5 and 60 minutes. If the “RepeatNo” is 0, this parameter has no effect.
GSM CONNECTION
After the correct settings are done, the SMS feature can be enabled by set “General Menu” - “Serial
& GSM Comm.” - “Type” to GSM.
Note: Power the GSM module before first attempt to connect the analyzer to GSM network.
If the PIN number is wrong, then it is not possible to make the cellular module connection. The
“GSM Wrong PIN Code” message is displayed.
If the first attempt to initialize GSM engine fails because of a wrong PIN, then no other attempt
will be allowed (prevention of sending three wrong PINs).
To make another attempt, a new PIN has to be entered and the analyzer has to be restarted.
The analyzer attempt to initialize the GSM module at startup.
GSM
If a wrong PIN was entered three times, the user has to extract the SIM card and manually enter the
PUK number using his own cellular phone to unlock the SIM card.
Note: There is no need to modify the default baud rate (1200) of the serial link when enable
the GSM feature. It will be switched to 9600 automatically.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments PCA310 and is the answer not in the manual?

Hanna Instruments PCA310 Specifications

General IconGeneral
BrandHanna Instruments
ModelPCA310
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals