EasyManuals Logo
Home>Hanna Instruments>Measuring Instruments>PCA310

Hanna Instruments PCA310 User Manual

Hanna Instruments PCA310
80 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #45 background imageLoading...
Page #45 background image
45
Max.Value:14.00 pH
Date:03/01/01 14:39
Min.Value:00.00 pH
Date:04/01/01 00:03
Clear Max. Value
Clear Min. Value
Note: It is not recommended to calibrate the analyzer at values below 2 mg/L in order to
maintain enough accuracy in the whole range. Calibration below 2 mg/L does not guarantee
declared accuracy outside an interval of ±50% from the calibration value.
Settings related to pH measurement are grouped in “pH Menu”. The following options are available:
MEASURE INFO
The analyzer calculates the maximum and minimum pH value since the first measurement.
The maximum and minimum can be quickly consulted on one pH measuring panel.
For more detailed information select “pH Menu” - “Measure Info” The “Max. Value” and “Min. Value”.
The time stamp when maximum and minimum occurs are also displayed in this menu.
To reset the maximum or minimum values, select the functions “pH Menu” - “Measure Info” - “Clear
Max. Value” or “pH Menu” - “Measure Info” - “Clear Min. Value”.
The maximum or minimum value is set to the current read value.
pH SETTINGS (PCA320, PCA330, PCA340)
• Press repeatedly ESC to exit from menu mode. The displayed chlorine concentration will be equal
with the calibration value.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments PCA310 and is the answer not in the manual?

Hanna Instruments PCA310 Specifications

General IconGeneral
BrandHanna Instruments
ModelPCA310
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals