ALARMS
Two alarm setpoints are available for chlorine: Alarm high and Alarm low.
The ALARM LED and alarm relay are activated when the chlorine concentration is higher than Alarm high
or lower than Alarm low.
To modify the alarm setpoints, enter the “Chlorine Menu” - “Alarms & Err Chlorine” and edit
Alarm Hi” or “Alarm Lo” value.
The alarms could be separately activated or inactivated.
To modify the alarms status, enter the “Chlorine Menu” - “Alarms & Err Chlorine” menu and edit
Alarm Hi” or “Alarm Lo” status. When the status is set to “Inactive”, the alarm is ignored.
Note: The Alarm high must be greater than Alarm low value. The analyzer display a warning
if the settings are incorrect.
The Alarm high setpoint and Alarm low setpoint could be quickly consulted on one chlorine
measuring panel. When an alarm is disabled, the —.— is displayed instead of alarm value.
CHLORINE SETTINGS