18
Filling Reservoir
Add distilled or low quality deionized water (mini-
mum 0.5 megohm/cm, maximum 1.5 megohm/
cm) to the reservoir up to the fill line. The MC8
reservoir holds approximately 3 liters; the MC10
reservoir, 4 liters. Do not overfill.
OPERATION
Caution
Use only distilled or low
quality deionized water
(minimum 0.5 megohm/cm,
maximum 1.5 megohm/cm).
Normal tap water contains
minerals that will corrode
sterilizer parts.
Note
Be sure reservoir drain valve is
closed before filling the reservoir.
Warning
Burn Hazard: Do not add water
to the reservoir while sterilization
cycle is in progress. Steam
vented into reservoir at end of
cycle could result in possible
burns.
The water in the reservoir could
be HOT if multiple cycles have
been run in succession.
Reservoir lid must be in place
when sterilizer is in use.
Figure 7
Filling Reservoir
Fill Funnel
Use Distilled
or Low Quality
Deionized Water
Only
Fill To Mark
Refill
Fill