35
2. Insert the drain fitting on the other end
of the plastic tubing into the drain port.
As soon as the drain fitting locks into
place, water will begin to exit the drain.
3. When the water flow stops, press the
tab on the drain port to release the
drain hose and fitting.
4. Store the drain tube and fitting.
5. Remove the reservoir lid and wipe the
inside of reservoir with a clean damp
cloth and mild disinfectant. Do not use
abrasive cleaners. Wipe the reservoir
clean of any disinfectant residue.
6. Remove and clean the reservoir filter.
7. Replace the reservoir lid and refill the
reservoir to the FILL mark in the fill
funnel with distilled or deionized water.
Cleaning Reservoir Filter
1. Unscrew the reservoir filter from its
fitting and rinse thoroughly with clean
water.
2. Scrub the outside of the filter with a
stiff-bristled brush (toothbrush) or
clean in an ultrasonic cleaner.
If the above cleaning method does not
effectively clean the filter, replace it.
OPERATOR MAINTENANCE
Hot Surface
Burn Hazard: The water may
be HOT if a cycle has been run
recently. Burns could occur if
hot water comes in contact with
skin.
Note
Reservoir cleaning is not required
as often on units with
non-recirculating feature.
Figure 25
Reservoir Filter Screen
To
Remove