Wipe all residue from the surfaces with a
dampened, lint-free cloth.
Cleaning and Disinfecting
Exterior Surfaces
The exterior surfaces may be cleaned and
disinfected with agents typically found in dental
and medical offices. Wipe the surface clean with
mild detergent and disinfect in accordance with
the instructions on the disinfectant agent
container.
Every 25 Cycles
Cleaning Reservoir
Draining Reservoir
A drain fitting and plastic tubing are provided to
drain the reservoir.
1. Insert the open end of the plastic tubing
into a container.
OPERATOR MAINTENANCE
Caution
Do not use steam to disinfect or
clean the sterilizer cabinet.
Extreme high temperatures
could cause plastic parts to
become distorted.
Caution
Do not use abrasive cleansers.
Hot Surface
Burn Hazard: Be sure sterilizer
is cool before cleaning
reservoir, The water in the
reservoir could be HOT if
sterilizer was in use just prior to
cleaning reservoir.
Note
Sterilizers equipped with an
optional non-recirculating system
may require less frequent
reservoir cleaning.
Figure 23
Draining Reservoir
Warning
Water May Be Hot
Drain
Fitting
Click In
To Drain
Release
Tab
Push To
Release
Drain
Port
Figure 24
Cleaning Reservoir
Clean Reservoir With
Disinfectant
Rinse Thoroughly
Before Filling With
Distilled Water