Daily
Cleaning Door Gasket
Wipe the door gasket and mating surfaces daily
with a clean damp cloth. Do not use abrasive
cleaners on the gasket or mating surface.
Periodically inspect the gasket for cracks or
damage. If replacement is necessary, refer to
Door Gasket Replacement.
Weekly
Cleaning Chamber, Trays and Rack
At least once a week, remove the trays and
rack from the chamber, and thoroughly clean
the inside of the chamber (especially the
bottom).
Clean the chamber, trays and rack with Harvey
brand chamber cleaner. Harvey Chamber
Cleaner works best in a dry chamber at room
temperature.
Operator Maintenance
Hot Surface
Burn Hazard: Be sure sterilizer is
cool before cleaning door gasket
and mating surface.
Hot Surface
Burn Hazard: Be sure sterilizer is
cool before cleaning chamber
surfaces and components.
Caution
Failure to keep the interior of the
stainless steel chamber free from
mineral deposits or other debris can
cause serious deterioration and
premature failure of the chamber.
Caution
Never use a wire brush, steel wool,
abrasive cleaners or cleaning
agents containing chloride. These
materials will damage the metal
surfaces.
Figure 22
Cleaning Chamber
Rub Area
To Be Cleaned
Rinse with
Water and
Wipe Clean
Figure 21
Door Gasket
Wipe With Clean,
Damp Cloth