Additional Information
Cancel Cycle
To cancel a sterilizing cycle or drying phase in
progress, press the STOP switch.
Emergency Power Off
To remove all power from the sterilizer, press
the OFF switch, then unplug the power cord.
Power Failure During Cycle
If a power failure occurs during a cycle, the
current program will be stored in memory.
When power is restored, the controls will
attempt to continue the cycle or will cancel the
cycle depending on the length of the power
interruption.
If you cannot start a cycle after power is
restored:
1. Unplug the unit.
2. Press and hold the OFF button while
plugging the unit back in.
3. Release the OFF button after
“PERFORMING SELF-TEST” is
displayed.
Warning
Volatile Reaction: Do not open
the door if there is a liquid load
in the chamber. The containers
may boil over or explode. Call for
authorized service.
Caution
The door will not open if there is
pressure or vacuum in the
chamber. Do not attempt to force
the door open. Press OFF, then
ON, to vent the chamber so that
the door can be opened.
Note
If a cycle is canceled
prematurely, reprocess the load.
Failure to do so may result in
non-sterile goods.