1. Activate the non-recirculating feature:
a. Press and release OFF switch,
then press again and hold. Do not
release.
b. Press and hold PROGRAM/SET
switch. Do not release.
c. Release the OFF switch ONLY.
d. Release the PROGRAM/SET
switch.
The sterilizer is now ready for you to proceed
with Step 2.
INSTALLATION
Warning
Non-Recirculating Option must be
installed before proceeding.
Note
This operation requires the use of
two hands. You must tell the
sterilizer that the non-recirculating
feature is ready. This only takes a
few seconds. If at any time you wish
to discontinue this activation mode,
press cycle stop.