32
PME-MP85
A08973.4 enHBM
Synchronization:
Synchronization is advisable for carrier-frequency fed transducers if
D the transducer cables of several devices are installed side by side
D the measuring points are close together and unshielded
Synchronisation prevents carrier frequency differences leading to disturbing
superpositions.
NOTE
When synchronizing several modules, one device has to be declared the
master. The remaining devices must be set to slave (use Setup).
Synchronization between modules should always even if you are working
without a CAN bus be effected with the ribbon cable.