7
Chapter 3 Instructions before installation
3.1 check before installation
Users please ensure whether the product you received is your purchase before
installation.
You had batter check all items according the packing list.
Whether the power site is close to the system.
Whether the feed water meets the requirements.
Whether the feed water pressure is normal.
Whether the feed water supply is nonstop.
Whether the feed water pipe has1/2″NPTF whorl connector.
Whether it is needed another frame for wall installation.
Whether the strength of the wall can bear the system (including tank full of water)
weight.
Whether the system’s height can clearly observe both LCD screen and keypad
operations on control panel.
Whether there is adequate space to substitute supplies and connecting pipe as well
around and behind the system.
Whether you have installed more than 1/2 ″ drains, not more than one meter far
away from the system, intended for discharging waste-water and water overflowing.
Whether the installment circumstances are consistent with the regulations.
Whether there is reserved space if you are going to install RS232 signal accessing
your computer.
Whether there is reserved space if you are looking forward to having moving handle
and pedal switch installed.
3.2 Option spare parts
3.2.1 Systematic data recording package CR-SP840
It includes systematic data recording software and RS232 connecting signal. Using it,
you can easily have SMART-MINI system and your computer connected, on which you can
install and record you using this recording system and water quality.
3.2.2 Enhanced pretreatment cartridge CR-SP102