6.3.7How to re-installation when the cartridges are not installed in the right
place.
The system will leak water when the cartridge installation has some mistakes. The
cartridges re-installation is needed. Refer to the following content.
Refer to this chapter and enter the maintenance setting, press the "Disp. / " or "Flush /
" keypad to select "consumable replacement". White on Black means that it has been
selected, press the "Stop /
" keypad to set.
Now you can remove the cartridge which has leak water, and re-install them according to
the instructions in previous chapter.
Press "Menu" keypad continuously to exit the maintenance setting, the system will
restart.
In the consumable reset process, you do not need to choose any type of the
cartridges after entering the consumable replacement interface. Do not press
“Stop/
keypad, otherwise it will cause the cartridge replace data recording errors.
6.4 System status check
There is an item for system status check in maintenance setting, which allows you to
understand the current system status.
Refer to this chapter into the maintenance settings, press the "Disp. / " or "Flush / "
button to select the "machine state" White on Black means that it has been selected. As shown
below.
System display
cartridge
replace
system
display
data
recorder
manufactory.
setting
Press "Stop / " keypad to confirm, LCD displays current working state. As shown
below.
FEED42s/cm
RO15μs/cm
DS96.5%
T25.4
PF036h
UP15MΩ.cm
UV0000h
MF0000h