34
Pull out the inlet/outlet water tube of the UV lamp, mark the connector to avoid mistakes
in the next installation process.
Unscrew the fixer of the UV lamp, as shown in figure A.
Remove the UV lamp hoop, as shown in B.
Remove the UV light from UV lamp holder out. Shown in C below.
Spin out the exhausted UV lamp from the stainless steel housing. As shown below.
Prepare the new UV light. Make sure that the new UV lamp in good condition without
damage.
Insert the new UV light into the stainless steel shell, and tighten.
Install the UV lamp on the holder, and install and tighten the fixer.
Insert the inlet/outlet water pipe into the corresponding connector.
Installed side door.
GO back to the control panel, press the "Stop /
" keypad to confirm the replacement
process is done.
After finishing the UV lamp replacement, you can press "Disp. /
▲
" or "Flush /
▼
"
keypad the selection other components replacement, or press "Menu" button to return to the
upper level of menu. Or you can continuously press "Menu" keypad to exit the maintenance
setting, the system will restart.
6.3.6 Micro-filter replacement
Enter the maintenance setting, press the "Disp. /
▲
" or "Flush /
▼
" keypad to select