39
replacement records.
Press "Menu" keypad to return to the previous menu, or press "Menu" keypad
continuously to exit the maintenance setting, the system restart.
6.7 Restore factory settings
Various parameter settings has been changed by users. When you need to return to the
default settings, you can choose this item.
Refer to Section 3.2 of this chapter into the maintenance settings. Press the "Disp. / ▲"
or "Flush /
▼
" keypad to select "Manu. setting" White on Black means that it has been
selected. As shown below. Press “Stop / ” keypad to confirm.
Manufactory setting
cartridge
replace
system
display
data
recorder
manufactory
setting
Load default?
car tr idg e
rep la ce
sys te m
dis pl ay
dat a
rec or der
man uf act or y
set ti ng
Press the "Stop / " keypad to confirm, the system restore the manufactory settings of
the parameter values.
Press "Menu" keypad to return to the previous menu, or press "Menu" keypad
continuously to exit the maintenance setting, the system reboots.
When the restart of system is complete, each parameter value is automatically set back to
manufactory setting.
Manufactory settings for all parameters are as follows:
Pretreatment cartridge alarm time: 400h
UV lamp alarm time: 4000h
Micro-filter cartridge alarm time: 6000h
Feed water alarm conductivity: 400μS/cm
Conductivity of pure water alarm: 20.0μS/cm
Reverse osmosis desalination rate alarm: 90%
Resistivity 0f ultrapure-water is 16.0 MΩ.cm
The display unit of is MΩ.cm
Volume of dispense setting: OFF
Obligate level setting: 60%
Dispensing flow is 0.5 l/m