99Using the HeartWare® Monitor
10
9
8
7
6
5
4
3
2
1
Appendix
Reference
Guides
Alarms and
Emergencies
Patient
Management
Surgical
Implant and
Explant
Monitor
Peripherals
and
Accessories
HVAD
®
PumpOverviewIntroduction
5.3 System Screens (continued)
Alarm Settings Tab (continued)
Figure 84: Permanently Silence [Controller Fault] Dialogue Box
The “Electrical Fault Audio” button appears during a medium priority [Electrical Fault] alarm
(Figure 85). The “Electrical Fault Audio” button can be used to permanently silence an [Electrical
Fault] alarm. However, the controller and monitor will continue to display the [Electrical Fault]
alarm until the condition resolves.
Figure 85: “Electrical Fault Audio” Button
Flow (L/min)
Time (s)
Logout
8
6
4
2
0
Speed/
Control
CONTROLLER FAULT
Call: ALARMS OFF
S
ee
/
n
r
l
Low Flow Alarm
Limit (L/min):
High Power Alarm
Limit (Watts):
Controller Fault
Audio:
Setup
Alarm
Settings
2
5.4
2500
5.0
L/min
RPM
Watts
Fixed
Sx Off
HW1234567
POD: 0
00:01:31
1
1.0
16.0
ON
Permanently silence
existing controller
fault alarm?
Yes
No
Flow (L/min)
Time (s)
Logout
8
6
4
2
0
Speed/
Control
ELECTRICAL FAULT
Call
S
ee
/
n
r
l
Low Flow Alarm
Limit (L/min):
High Power Alarm
Limit (Watts):
Electrical Fault
Audio:
Setup
Alarm
Settings
2
5.4
2500
5.0
L/min
RPM
Watts
Fixed
Sx Off
HW1234567
POD: 0
00:01:31
1
1.0
16.0
ON
For additional information on alarms, see Section 8.1.
WARNING! ALWAYS investigate, and if possible, correct the cause of any alarm. Silencing an alarm does
not resolve the alarm condition.