2. Press the
[START]
button.
🡆 The centrifugation run is started.
The
[START]
button ashes until the rotor is read in.
The
[START]
button lights up during the centrifugation run.
The rotor speed or the resulting RCF value, the temperature in
the centrifuging chamber (only for centrifuges with cooling) and the
remaining time are displayed during the centrifugation run.
3. Ramp-down takes place with the selected ramp-down parameter
after the time has elapsed or if the centrifugation run is cancelled.
🡆 The ramp-down parameter is displayed.
The right side of the
[STOP/OPEN]
button lights up when the
centrifuge is in ramp-down.
The left side of the
[STOP/OPEN]
button lights up when the rotor is
at a standstill.
The light on the
[START]
button and the right side of the
[STOP/
OPEN]
button go out.
6.7.3
Short-term centrifugation
Personnel:
■
Trained user
1. Press and hold the
[START]
button.
🡆 The
[START]
button ashes until the rotor is read in.
The
[START]
button lights up during the centrifugation run.
Timing starts at 00:00.
The rotor speed or the resulting RCF value, the temperature in
the centrifuging chamber (only for centrifuges with cooling) and the
elapsed time are displayed during the centrifugation run.
2. Release the
[START]
button to end the centrifugation run.
🡆 The ramp-down parameter is displayed.
The right side of the
[STOP/OPEN]
button lights up when the
centrifuge is in ramp-down.
The left side of the
[STOP/OPEN]
button lights up when the rotor is
at a standstill.
The light on the
[START]
button and the right side of the
[STOP/
OPEN]
button go out.
6.7.4
Changing settings during centrifugation
It is not possible to change settings during centrifugation when working with
program links or when a program lock has been set.
The runtime, speed, relative centrifugal force (RCF), ramp-up and ramp-
down parameters and the temperature (only for device with cooling) can be
changed during centrifugation.