■ Rotor information
■
Centrifuge program version
■ Program version for the frequency inverter
The rotor is stationary.
1. Press and hold the
[PROG]
button.
🡆
‘***Machine Menu***’
is displayed after 8 seconds.
2. Press the
[PROG]
button repeatedly until
‘-> Info’
is displayed.
3. Press the
[START]
button.
🡆
The centrifuge model is displayed.
4. Press the
[PROG]
button.
🡆 The mains voltage is displayed
5. Press the
[PROG]
button.
🡆 The rotor code (Rotor), the maximum rotor speed (Nmax) and the
centrifuging radius (R) of the last rotor recognised by the rotor
detection are displayed.
The last rotor recognised is marked with an asterisk (*).
The
[Rotary knob]
can be used to display the information of the
rotors approved for the centrifuge.
6.
Press the
[PROG]
button.
🡆
The program version of the centrifuge is displayed.
7.
Press the
[PROG]
button.
🡆
The program version of the frequency converter is displayed.
8.
Press the
[STOP/OPEN]
button twice to exit the
‘-> Info’
menu
or
Press the
[STOP/OPEN]
button three times to exit the
‘***Machine
Menu***’
.
7.6.1.1 Centrifuge address
The address of the centrifuge is set to ]=29 ex works. Address set.
7.6.2 Cycle counter
The centrifuge is equipped with a cycle counter. The cycle counter counts
the run cycles (centrifugation cycles) of the various rotor codes.
For swing-out rotors, the cycle counter is used to record the run cycles
(centrifugation runs) of the buckets.
If the rotor is detected for the rst time by the rotor detection, the centrifu-
gation run is cancelled.
‘Enter max cycles = á30000 ñ’
is displayed after
pressing any button. The maximum permissible number of run cycles indi-
cated on the bucket must be entered before the centrifugation run can be
restarted.
The cycle counter can be disabled for rotors and buckets that are not
marked with the maximum permissible number of run cycles. The number
of run cycles (centrifugation runs) of the rotor code used is displayed briey
every time the lid is opened.
If the maximum permissible number of bucket run cycles entered is
exceeded,
‘*MAX CYCLES PASSED*’
is displayed after each start of a
centrifugation run.
The centrifugation run must be restarted. The buckets must be replaced
with new ones.