5. Press the
[PROG]
button.
🡆 The set RCF value is saved.
7.1.8 Centrifugation of substances or mixtures of substances with a density higher than
1.2 kg/dm
3
The density of the substances or mixtures of substances must not exceed
1.2kg/dm³ during centrifugation at maximum speed. The speed must be
reduced for substances or substance mixtures with a higher density. The
permissible speed can be calculated using the following formula:
For example: Maximum speed 4000 RPM, density 1.6kg/dm³
If, in exceptional cases, the maximum load indicated on the bucket is
exceeded, the speed must also be reduced. The permissible speed can
be calculated using the following formula:
For example: Maximum speed 4000RPM, maximum load 300g, actual
load 350g
Please contact the manufacturer if you are not sure.
7.2
Programming
7.2.1 Write protection for programs
Write protection can be enabled or disabled when the rotor is at a standstill.
1. Open the desired program.
2. Press the
[PROG]
button.
🡆 The RCL parameter is displayed.
3. Press and hold the
[PROG]
button.
🡆 The STO parameter is displayed.
‘Set Protection = 1-’
is displayed after 8 seconds.
4. Use
[Rotary knob]
to set
‘+’
or
‘-’
.
+ = The program is write-protected
- = The program is not write-protected
5. Press the
[START]
button.
🡆 The setting is stored.
7.2.2
Opening or loading programs
1. Press the
[PROG]
button.
🡆 The RCL parameter is displayed.