AB4701en_SARev.: 05 / 11.2023
40 / 62
Software operation
enable / disable
4. Press the
[PROG]
button repeatedly until
‘Error 58 Temp 15°C’
is
displayed.
5. Use the
[Rotary knob]
to set the desired value.
Adjustable from 4°C to 25°C, in 1°C increments, as well as the "dis-
abled" setting. Temperature monitoring is deactivated when "disabled"
is set.
6. Press the
[START]
button.
🡆 The setting is stored.
‘Store Settings...’
is displayed briey.
‘-> Settings’
is then displayed.
7. Press the
[STOP/OPEN]
button once to exit the
‘Settings menu’
or
Press the
[STOP/OPEN]
button twice to exit the
‘***Machine Menu***’
.
7.5
Heating (for centrifuges with heating)
During the centrifugation run, the centrifuging chamber is heated to the
preselected temperature if required. The heating is switched off when the
rotor is at a standstill.
Swing-out rotors and angle rotors must run at maximum speed.
CAUTION
Danger of burns from hot surfaces.
The surface temperature of the heating element in the centri-
fuging chamber can be up to 500°Cor 932°F.
Do not touch the heating element.
NOTICE
Damage to plastic buckets due to excessive temperature
Plastic buckets may only be used at temperatures up to
a maximum of 40°C or 104°F.
The rotor is stationary.
1.
Press the
[T/°C]
button repeatedly until
‘Heater = off’
or
‘Heater = on’
is displayed.
2.
Use
[Rotary knob]
to set
‘off’
or
‘on’
.
off = heating disabled
on = heating enabled
3.
Press the
[T/°C]
button or the
[START]
button.
🡆
The settings are stored.
Centrifugation data is displayed.
7.6 Machine Menu
7.6.1 Querying system information
The following system information can be queried:
Centrifuge model
Mains voltage