3. Press the
[START]
button.
🡆
‘Multi program recall...’
is displayed briey.
The centrifugation data of the rst program of the program link is
displayed, along with the total runtime of the program link.
7.7.3 Enabling or disabling program links
1. Press and hold the
[PROG]
button.
🡆
‘***Machine Menu***’
is displayed after 8 seconds.
2. Press the
[PROG]
button repeatedly until
‘-> Settings’
is displayed.
3. Press the
[START]
button.
🡆
‘SOUND / BELL = off’
or
‘SOUND / BELL = on’
is displayed.
4. Press the
[PROG]
button repeatedly until
‘ Multi programs = off’
or
‘ Multi programs = on’
is displayed.
5. Use
[Rotary knob]
to set
‘off’
or
‘on’
.
off = program link disabled
on = program link enabled
6. Press the
[START]
button.
🡆 The setting is stored.
‘Store Settings...’
is displayed briey.
‘-> Settings’
is then displayed.
7. Press the
[STOP/OPEN]
button once to exit the
‘Settings menu’
or
Press the
[STOP/OPEN]
button twice to exit the
‘Machine Menu’
.
8 Cleaning and care
8.1 Overview table
Chap. Task to execute
8 Cleaning and care 51
8.3 Cleaning 52
8.3 Cleaning the device X 52
8.3 Cleaning the biosafety systems X 53
8.3 Cleaning the accessories X 53
8.4 Disinfection 53
8.4 Disinfecting the device X 53
8.4 Disinfecting the accessories X 54
8.5 Maintenance 54