43 / 62
Rev.: 05 / 11.2023
AB4701en_SA
Software operation
Disabling the cycle counter
4. Press the
[PROG]
button repeatedly until
‘Cycles = disabled’
is dis-
played when the cycle counter is disabled.
If run cycles are displayed, the cycle counter is already enabled.
5. Press the
[RCF]
button repeatedly until the maximum permissible
number of run cycles is displayed in brackets áñ.
6. Use the
[Rotary knob]
to set the maximum permissible number of run
cycles indicated on the bucket.
7. Press the
[START]
button.
🡆 The settings are stored.
‘Store cycles ...’
is displayed briey.
The run cycles are displayed.
8. Press the
[OPEN/STOP]
button twice to exit the
‘Operating Time’
menu
or
Press the
[OPEN/STOP]
button three times to exit the
‘Machine
Menu’
.
The rotor is stationary.
1.
Press and hold the
[PROG]
button.
🡆
‘***Machine Menu***’
is displayed after 8 seconds.
2. Press the
[PROG]
button repeatedly until
‘-> Operating Time’
is dis-
played.
3. Press the
[START]
button.
🡆 The external operating hours are displayed.
4. Press the
[PROG]
button repeatedly until, with the cycle counter ena-
bled, the run cycles are displayed.
If
‘Cycles = disabled’
is displayed, the cycle counter is already disa-
bled.
5. Press the
[RCF]
button repeatedly until the maximum permissible
number of run cycles is displayed in brackets áñ.
6. Use
[Rotary knob]
to set the maximum permissible number of run
cycles to
‘0’
.
7. Press the
[START]
button.
🡆 The settings are stored.
‘Store cycles ...’
is displayed briey.
‘Cycles = disabled’
is displayed.
8. Press the
[OPEN/STOP]
button twice to exit the
‘Operating Time’
menu
or
Press the
[OPEN/STOP]
button three times to exit the
‘Machine
Menu’
.
7.6.3
Querying operating hours, centrifugation runs and cycle counter
The operating hours are divided into internal and external operating hours.
Internal operating hours
‘OP Time int =’
: Total time for which the device
has been switched on.
External operating hours
‘OP Time ext =’
: Total time of centrifugation
runs to date.