7. Press the
[STOP/OPEN]
button once to exit the
‘Settings menu’
or
Press the
[STOP/OPEN]
button twice to exit the
‘Machine Menu’
.
7.6.5
Enabling or disabling ramp-up and ramp-down times
The rotor is stationary.
1.
Press and hold the
[PROG]
button.
🡆
‘***Machine Menu***’
is displayed after 8 seconds.
2. Press the
[PROG]
button repeatedly until
‘-> Settings’
is displayed.
3. Press the
[START]
button.
🡆
‘SOUND / BELL = on’
or
‘SOUND / BELL = off’
is displayed.
4. Press the
[PROG]
button repeatedly until
‘Ramp Unit = Steps’
or
‘Ramp Unit = Steps / Time’
is displayed.
5. Use
[Rotary knob]
to set
‘Steps’
or
‘Steps / Time’
.
Steps = ramp-up and ramp-down times disabled,
Steps / Time = ramp-up and ramp-down times enabled.
6.
Press the
[START]
button.
🡆 The setting is stored.
‘Store Settings...’
is displayed briey.
‘-> Settings’
is then displayed.
7. Press the
[STOP/OPEN]
button once to exit the
‘Settings menu’
or
Press the
[STOP/OPEN]
button twice to exit the
‘Machine Menu’
.
7.6.6
Program lock
The following program locks can be set when the rotor is at a standstill:
LOCK 1 LOCK 1 is displayed.
Programs can only be retrieved, not changed.
LOCK 2 LOCK 2 is displayed.
No programs can be retrieved or changed. The centrifuge
can be controlled via the interface (only for centrifuges
with an interface).
LOCK 3 no status indicator
No program lock. Programs can be retrieved and
changed.
1.
Press and hold the
[PROG]
button.
🡆
‘***Machine Menu***’
is displayed after 8 seconds.
2.
Press the
[PROG]
button repeatedly until
‘-> Change Lock’
is dis-
played.
3.
Press the
[START]
button.
🡆 Lock status is displayed.
If no PIN is entered, e.g.
‘LOCK = á3ñ conrm by START’
is dis-
played.
If a PIN is entered, e.g.
‘LOCK = 3’
is displayed.