Function Checks
Chapter 2: Troubleshooting Procedures
Page 2 - 4 Electrically Operated General Ward Bed
Operation and Maintenance Manual (154588 REV 1)
9. Press the Knee Up control.
The knee section rises to the high position without stopping or the audible
alarm sounding.
Yes No
Go to RAP 2.3.
10. Press the Knee Down control.
The knee section goes down to the low position without stopping or the
audible alarm sounding.
Yes No
Go to RAP 2.3.
11. Press the Head Up control.
The head section rises to the high position without stopping or the audible
alarm sounding.
Yes No
Go to RAP 2.1.
12. Press the Head Down control.
The head section goes down to the low position without stopping or the
audible alarm sounding. (Do not consider the knee section motor.)
Yes No
Go to RAP 2.5.
13. Press the Head Down control.
The head section goes down, then the knee section continues to drop when
the head section reaches the low position.
Yes No
Go to RAP 2.5.
14. Unplug the power from the power source and test the functions checked in
step 12 to step 14.
The bed functions operate correctly.
Yes No
Go to RAP 2.6.
15. Press the Head Up control until the section reaches the high position.
NOTE:
Someone should be on the bed so that this test can be carried out conclusively.
16. Pull one of the CPR control handles.