Function Checks
Chapter 2: Troubleshooting Procedures
Electrically Operated General Ward Bed Page 2 - 3
Operation and Maintenance Manual (154588 REV 1)
2
Function Checks
1. Initial Actions have been performed.
Yes No
Go to “Initial Actions” on page 2-2.
2. Set the brakes, raise and lock the siderails in the high position. Plug the
power cord into an applicable power source.
3. Operate the Trendelenburg function.
This function works.
Yes No
Go to RAP 2.7.
4. Lock at least one bed function (such as the head section).
The corresponding indicator comes on.
Yes No
Go to RAP 2.2.
5. Lock out all bed functions.
All of the lock out indicators come on (except the bed not down indicator).
Yes No
Go to RAP 2.2.
6. Unlock all of the functions.
All of the lock out indicators go off (except the bed not down indicator).
Yes No
Go to RAP 2.2.
7. Press the Bed Up control.
The bed rises to the high position without stopping or an audible alarm
sounding.
Yes No
Go to RAP 2.4.
8. Press the Bed Down control.
The bed goes down to the low position without stopping or an audible
alarm sounding. In addition, the bed not down indicator goes off when the
bed reaches the lowest position.
Yes No
Go to RAP 2.4.