2.5 Backrest Malfunction
Chapter 2: Troubleshooting Procedures
Electrically Operated General Ward Bed Page 2 - 13
Operation and Maintenance Manual (154588 REV 1)
2
2.5 Backrest Malfunction
1. Does at least one of the other functions operate correctly?
Yes No
Go to RAP 2.1 on page 2-7.
2. Is the backrest lockout function unlocked?
Yes No
Unlock the backrest function.
3. Is the handset plugged into the socket correctly?
Yes No
Connect the handset.
4. Is the ACP connected to the control box?
Yes No
Connect the ACP to the control box.
5. Test the actuator. Trace the wire from the actuator to the control box, and
label it. Trace the wire from a working actuator to the control box and label
it. Switch the two connections and use the handset to actuate the known
good actuator.
The backrest actuator fails to extend/retract when the known good function
is activated.
Yes No
Switch the connections back to original, and then replace the
backrest actuator. Go to “Function Checks” on page 2-3.
6. Without switching the wiring back to original positions, swap the ACP
with a known good ACP. Use the handset to activate the known good
function. The backrest actuator fails to extend/retract when the known
good function is activated.
Yes No
Switch the connections back to original, and then replace the
backrest actuator. Go to “Function Checks” on page 2-3.