2.7 Trendelenburg Malfunction
Chapter 2: Troubleshooting Procedures
Page 2 - 16 Electrically Operated General Ward Bed
Operation and Maintenance Manual (154588 REV 1)
2.7 Trendelenburg Malfunction
1. Does the HI/LO function operate correctly?
Yes No
Go to RAP 2.1 on page 2-7.
2. Is the trendelenburg lockout function unlocked?
Yes No
Unlock the function.
3. Is the handset plugged into the socket correctly?
Yes No
Connect the handset.
4. Is the ACP connected to the control box?
Yes No
Correctly connect the ACP to the control box.
5. Replace the control box with a known good control box. The trendelenburg
continues to function incorrectly.
Yes No
Replace the affected control box. Go to “Function Checks” on
page 2-3.
6. Contact Medicraft Hill-Rom Customer Service.