Specifications
98 Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Electromagnetic Immunity Guidance
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The P7800 is intended for use in the electromagnetic environment specified below. The customer or the user of the
P7800 should make sure it is used in such an environment.
Immunity Test
EN & IEC
60601-1-2
(4th Ed.)
Test Level
Compliance
Level
Electromagnetic Environment—Guidance
Conducted
RF-Immunity
IEC 61000-4-6
3 V (80% AM)
150 KHz to
80 MHz
(6 V in ISM
Bands per
CISPR-11)
3 V (80% AM)
150 KHz to
80 MHz
(6 V in ISM
Bands per
CISPR-11)
Field strengths from fixed RF transmitters, as determined by
an electromagnetic site survey, should be less than the
compliance level in each frequency range.
Radiated
RF-Immunity
IEC 61000-4-3
3 V/m
80 MHz to
2700 MHz
10 V/m
80 MHz to
2700 MHz
Interference may occur in the vicinity of equipment marked
with this symbol.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Note 3: Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast can not be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the P7800 is used exceeds the applicable
RF compliance level above, the P7800 should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the P7800.