Standard Features
22 Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Use the CPR Controls
WARNING:
Warning—Failure to start the deflation of the surface before you start CPR could cause CPR to be
ineffective.
To Lower the Head of the Bed
Pull and hold the red CPR handle until the
head section is in a flat position. You must
hold the CPR handle until the head section
reaches the flat position. If you release the
handle during its operation, the head section
will stop lowering. The rate of descent
depends on the weight of the patient, but on
average it takes approximately 5 to 10
seconds for the head section to lower.
To Deflate the Air Surface
Turn the air surface CPR mechanism
clockwise until it stops and you hear air
being released. The surface will start to
deflate.
NOTES:
When the air surface is completely
deflated, the bed's sleep deck can be
used as a backboard.
The headboard and/or patient helper can
be removed for patient access for patient
intubation or to insert a central line.
To Inflate the Air Surface after CPR
1. Turn the air surface CPR mechanism counter-clockwise until it locks into position.
2. On the air supply unit, press Max Inflate to quickly inflate the surface.
3. After the surface is fully inflated, press Max Inflate again to toggle it off.
NOTE:
If the CPR mechanism is not fully closed when the unit is in Max Inflate mode, an alert will sound to let
you know of the air loss through the CPR valve.
POWER CORDS
There are two power cords:
:
The power cord on the patient-right side of the bed
supplies power to the bed frame articulation controls
and charges the built-in battery back-up and
IntelliDrive® XL Transport System batteries.
North America