Single Mode Bed Exit Alert System
50 Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
TO DEACTIVATE THE BED EXIT ALERT SYSTEM
TO ADJUST THE ALERT VOLUME
TO CHANGE THE ALERT TONE
NOTE:
We recommend that you use the same tone on all beds of a particular unit or floor, and do not change
the tone without facility authorization.
1. Activate one of Bed Exit modes. It is recommended to use another caregiver instead of a patient to
activate the Bed Exit mode.
NOTE:
To activate a Bed Exit mode, a minimum of 113 kg (250 lb) must be on the bed.
2. Activate the alert by having the caregiver exit the bed.
3. Press and hold the Volume control.
4. While you press the Volume control, press the Out-of-Bed control.
5. Press and release the Out-of-Bed control until the desired tone is reached.
6. Clear the alert condition.
Press the Enable key until its indicator is on solid, and then press Out-of-Bed control
until the indicator goes off.
1. The patient must be on the bed.
2. The system must be armed.
3. Press the Enable key until its indicator is on solid.
4. Press and release the Volume control until the indicator that is next to the
desired volume setting comes on.