Preventive Maintenance
84 Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Bed Frame
Replace the batteries if any of these conditions occur (refer to the Compella™ Bariatric Bed System Service
Manual):
The battery indicator does not come on within 3 minutes of bed connection to AC mains.
The battery indicator does not stop flashing (low condition) within 12 hours of bed connection
to AC mains.
Successive transports of 4 hours or less cause the batteries to discharge to low condition as
indicated by a flashing battery indicator.
IntelliDrive® XL Transport System
Contact your facility-authorized maintenance person or Hill-Rom Technical Support if the transport
system automatically shuts down power before the final battery charge indicator flashes. The batteries
will need to be replaced.
After the batteries are replaced, charge them for a minimum of 20 hours before use.
AIR SUPPLY UNIT—FILTERS REPLACEMENT
WARNING:
Warning—Failure to clean the filters may cause damage to the air supply unit which could prevent the
unit from supplying the correct support surface pressures. Patient injury could occur.
Excessive lint, dust, and/or smoke can clog the
filter. To keep the air supply unit operating at its
best, good filter maintenance is critical. After 800
hours of operation, when the air supply unit is
turned on, the Replace Filter screen shows as a
reminder to replace the filters. To replace the filters
later, press Remind me later. When the air supply unit is turned on again, the Replace Filter screen will
show.
For filter replacement and cleaning, see “Clean the Air Supply Unit Filters on page 82.