Cleaning and Disinfecting
78 Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Syringe-Driver Holder (AC963)
To help prevent personal injury and/or equipment damage, follow these warnings and cautions:
• Warning—Do not exceed the load capacity of the syringe-driver holder; refer to the value
shown on the holder.
• Warning—Do not install the holder so that it is towards the bed. To do so could cause
interference with the bed and siderail articulations.
The holder is designed to hold syringe-drivers and is to be installed in the equipment sockets on either
side of the head end of the bed.
To Adjust the Holder Position
1. Hold the table as you loosen the knob.
2. Adjust the table as applicable, and then tighten the knob.
CLEANING AND DISINFECTING
To help prevent injury and/or equipment damage, obey these warnings:
• Warning—The potential for electrical shock exists with electrical equipment. Failure to follow
facility protocol could cause death or serious injury.
• Warning—Do not reuse wiping material for multiple steps or on multiple products.
• Warning—Harmful cleaning solutions may cause skin rash and/or irritation upon contact.
Follow the manufacturer’s instructions found on the product label and Safety Data Sheet (SDS).
• Warning—Lift and move items correctly. Do not twist, and seek assistance when necessary.
Make sure the bed is at a correct height to lift items off the bed.
• Warning—Fluid spills on to the bed electronics could cause a hazard. If such a spill occurs,
unplug the bed and remove it from service. When fluid spills occur outside of what is seen in
normal use, immediately do as follows:
a. Unplug the bed from its power source.
b. Remove the patient from the bed.
c. Clean the fluid spill from the bed system.
d. Have maintenance examine the system completely.
e. Do not put the bed back into service until it is completely dry, tested, and found to be safe
to operate.