Specifications
100 Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Recommended Separation Distances between Portable and Mobile RF
Communications Equipment and the P7800 Model
The P7800 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the P7800 can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the P7800 as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter,
W
Separation distance according to frequency of transmitter, m
150 kHz to 80 MHz
d = 1.2 P
80 MHz to 800 MHz
d = 1.2 P
800 MHz to 2.5 GHz
d = 2.33
P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and