Measurement Method
6.6
V10 Flicker
97
2.
Making 3196 settings (measurement settings)
MEASURE
Make the following settings.
Setting Item Selections
123ch/ Wiring
3P3W3M or 1P2W
123ch/ U Range
150V
123ch/ PT Ratio
60
UReference
110V
Frequency
50 Hz
PLL Source
U1
U CalcType
LINE-LINE (3P3W3M) or
PHASE-N
Flicker
V10
TimePlot U
U
Flicker CH
U1
EN50160
OFF
Select item
Select from
pull-down
menu
Confirm
3.
Connecting to the lines to be measured
Use the same method as with three-phase checking. (page 59)
With single phase 2-wire connection, connect only the CH1 voltage.
4.
Checking the connections
Use the same method as with three-phase checking. (page 60)
With single phase 2-wire connection, connect only the CH1 voltage.