6.6
V10 Flicker
100
Quick Start Manual
(2) To check how V10 flicker is varying
GRAPH
You can only record the channel that is selected as the flicker mea-
surement channel.
The vertical and horizontal scales of the fluctuation graph change auto
-
matically during recording.
If you want to change the vertical scale, you can do so while recording
is in progress; however, you must first stop recording if you want to
change the horizontal scale.
The
V10 flicker value is calculated once a minute without gaps,
regardless of the set interval time.
You can confirm the contents of settings (SYSTEM-DF2) and wave-
form display (VIEW-DF1/2/3/4) during recording, but you cannot
change them.
When you want to change the settings, you must stop recording first.
Recording start time
Current time
Shows the V10 values
every 1 minutes.
8.
Ending the recording interval
Use the same method as when measuring power fluctuation. (page 68)