EasyManua.ls Logo

Hioki 3196 - Page 75

Hioki 3196
158 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1
2
3
4
5
6
7
8
9
10
11
Measurement Method
6.2 Change in Power
69
9.
Viewing measured data that has been recorded
Viewing the change in MAX, MIN, and AVE values for RMS
voltage over a 5-day interval
RMS
This indicates the
area that displays
the fluctuation graph
for recorded internal
memory capacity.
To display the fluctuation graph for active power, reactive power,
frequency, and voltage
To change the scale for the Y- or X-axis
Use the same method as three-phase checking for cursor mea-
surement and scrolling.
To measure the cursor(page 62)
To scroll the waveform(page 62)
Displayed items: P, Q, Freq, U, Uave,
Displayed channel: CH1, CH2, CH3, sum
(Shows 1, 2, 3: channel, sum: total value for three-phase,
and ave: average value for three-phase.)
Y-axis Y div: x1, x2, x5, x10, x25, x50
X-axis T div: x1, x1/2, x1/4, x1/8, x1/16, x1/32, x1/64
MAX value: Yellow
AVE value: Green
MIN value: Red
Recording start time
Present tim
e
You can display either screen 1 or screen 2.

Table of Contents

Other manuals for Hioki 3196

Related product manuals