Measurement Method
6.3 Harmonic Fluctuation
75
8.
Ending the recording interval
Use the same method as when measuring power fluctuation. (page 68)
9.
Viewing recorded measurement data
Monitoring fluctuation in MAX, MIN, and AVE values for har-
monic RMS voltage over a 1-week period
(1) Checking whether total voltage distortion specified in the IEC
Standards has been exceeded.
Checking using the time series graph
RMS
To display the fluctuation graph for the total voltage distortion facto
r
Total voltage distortion: U-THD (Select from CH1, CH2, and
CH3.)
Recording start time
Current tim
e
This indicates the
area in recorded in-
ternal memory that is
shown in the fluctua-
tion graph.
MAX value: Yellow
AVE value: Green
MIN value: Red
An event occurs when the total voltage distortion factor or harmonic
current limit specified in the IEC Standards is exceeded.
If the mark is displayed in the time series graph, an event has
occurred. However, an event occurs whenever recording starts and
stops.