159
Functional Specications
Comparator Operation Judgment:
HIGH judgment: Measured value > upper limit value
IN judgment: Upper limit value ≥ measured value ≥ lower
limit value
LOW judgment: Lower limit value > measured value
Judgmentdelay:
Outputs judgment results once the same judgment has been
made the set number of times. This setting is valid only when
auto-hold operation is disabled and the instrument is in the
RUN state.
Settings Comparator: ON, OFF (BIN: forced OFF)
Upper limit value and lower limit value: -1000 V to +1000 V
(when the scaling function is set to ON, -1000 GV to 1000
GV), ON, OFF
(IN judgment when both the upper and lower limit values are
set to OFF.)
Number of setting digits: 7
Judgment delay: ON, OFF
Judgment delay count: 2 to 10
Judgment tone: OFF, TYPE1, TYPE2, TYPE3
Number of beeps: 1 to 5, continuous
Default settings Comparator: OFF
Upper limit value and lower limit value: 0 V, ON
Judgment delay: OFF, 2
HIGH judgment tone: OFF
IN judgment tone: OFF
LOW judgment tone: OFF
Number of beeps: 2
BIN Judgment BIN nos. 0 to 9 (Out of BINs)
IN judgment: Upper limit value ≥ measured value ≥ lower
limit value
OUT judgment: Lower limit value > measured value,
measured value > upper limit value
Settings BIN: ON, OFF (COMP: forced OFF)
Upper limit value and lower limit value: -1000 V to +1000 V
(when the scaling function is set to ON, -1000 GV to +1000
GV)
Number of setting digits: 7
Default settings BIN: OFF
Upper limit value and lower limit value: 0 V
Absolutevaluejudgment
Operation Performs comparator judgment or BIN judgment while
ignoring the sign of the measured value.
Settings Absolute value judgment: ON, OFF
Default settings Absolute value judgment: OFF
Auto hold Operation Automatically holds the measured value when it falls within
the hold range.
Measurement settings are xed as follows:
Integration time: MEDIUM; input resistance: 10 M
Ω
Continuous measurement: RUN; contact check: ON
Settings Auto hold: ON, OFF
Hold range: 0.001% of the range to 1.000% of the range
Default settings Auto hold: OFF
Hold range: 0.1% of range
13
Specications
w ww . . co m
information@itm.com1.800.561.8187