40
Starting Measurement
3
1
2
Setting the measurement count
˄ Increases the setting by 1.
˅ Decreases the setting by 1.
(Default setting: 1, valid setting range: 1 to 5000
measurements)
One contact check and delay will be inserted after trigger input. Measurement will then continue
without any delay until the next trigger input.
Self-calibration will not be performed until the set number of measurements is performed. In the
event that the value obtained by multiplying the integration time by the measurement count is
greater than 1 minute, manage the ambient temperature such that it does not vary more than
±1°C. (For an example, see “Appx. 5 Self-calibration” (p. Appx.11))
For more information: “Setting the integration time” (p. 67), “6.3 Contact Check” (p. 71)